Choksi Laboratories

Home» Calibration Services» Medical Device Calibrations

Medical Device Calibrations

Calibrating medical equipment is crucial for accurate healthcare. It ensures precise diagnosis, monitoring, therapy, and treatment. Proper calibration prevents misdiagnoses caused by poorly adjusted machines. For human health, absolute accuracy is essential, making standard calibration of hospital systems imperative. This process aims to provide advanced and error-free healthcare for the well-being of the population. We, Choksi Lab, ensure the service to the medical sector with high precise and trust. We also provide engraving (permanent marking id number) on volumetric apparatus on customer request.

We provide services for various instruments, including but not limited to:

Scope of Services:

Instruments Range Method for calibration
Autoclave (Pressure Temperature Time)
-0.87 bar – 600 bar
Using Digital Dead weight/ Pressure Gauge, 6 ½ digit multimeter and Pneumatic Pressure Pump by comparison method
Oxygen pressure gauge
0 – 300 bar
Using oxygen cylinder and pressure gauge by comparison method
Centrifuge machine
30 rpm – 25000 rpm
Using Digital Tachometer by comparison method
BP apparatus
0- 300 mmHg
Using digital manometer by comparison method
Nebulizer
0- 100 mbar
Using digital manometer by comparison method
Trans illuminator light source
100 – 10000 lux
Using lux meter comparison method
Radiant Warmer
30 °C – 50 °C
Using Standard RTDs and temperature indicator by Comparison Method
Micro Balance (d= 0.001mg), Class I and coarser
0.05mg – 20 g
Using Standard Weights of E1 and E2 class accuracy
Weighing scales (baby/ store)
0.5 mg – 50 kg
Using Standard Weights (E1 class – F1 class)
Electronic Tourniquet (time)
0- 24 hr
Comparison by Stop watch
Boyels apparatus
0-20 bar
Using pressure gauge and pump by comparison method
OT light
100 – 10000 lux
Using lux meter comparison method
Blade (measurement)
0- 25 mm
Using profile projector & micrometer (Least count 0.001 mm) (Angle: 1minute)
Freezer
-80 °C – 200 °C
Using SPRT with Thermometer, Dry Well Calibrator by Comparison Method
Electrode pH meter
0- 50 mV
Using multi function calibrator by comparison method
Scroll to Top